Tuttnauer Valueklave 1730 Autoclave
Defect: Water enters chamber after unit is exhausted and the door is closed
Water level is above open end or there is a hole in the condensation coil creating a vacuum.
Reduce water level to 1″ below safety valve and open end of condensation coil must be above water level. Replace condensation coil if necessary.