Service and repair for your

Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave

Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave
Most common repairs + troubleshooting for the Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave at Acme:
  1. Power Interruption (E001): Occurs when there's an unexpected loss of power, possibly due to faulty outlets or overheating issues.
  2. Stop Button Pressed (E002): Indicates the stop button was pressed during a cycle, potentially pointing to a faulty display board.
  3. On/Standby Button Pressed (E003): This error is triggered if the On/Standby button is accidentally engaged during operation.
  4. Door Ajar (E004): Signals the door wasn't properly latched, which might involve a faulty door switch or alignment issue.
  5. High Pressure (E005): When the pressure exceeds 35 PSI during a cycle, suggesting overloading or a steam trap issue.
  6. High Internal Temperature (E006): Occurs when internal temperatures exceed 277°F, indicating potential overloading or sensor issues.
  7. Low Pressure (E007): Indicates a potential steam leak, requiring checks on the door gasket or internal components like valves.
  8. Door Not Opening (E008): Occurs at the cycle's end, preventing transition to the dry cycle, possibly due to worn-out door mechanisms.
  9. Pressure Buildup During Dry Cycle (E009): Suggests issues completing the dry cycle, potentially due to calibration errors.
  10. Software Issues (E010, E011, E012): General software problems that may necessitate a reboot or PC board replacement.

est repair cost: $45 per hour

387 parts + components available at: ACME REVIVAL DENVER

  • Exterior

    confirmed fit for: Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave

  • Internal (Mechanical)

    confirmed fit for: Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave

  • Circuitry

    confirmed fit for: Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave

  • Plastics

    confirmed fit for: Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave

$45 Per Hour

Acme bills $45 per hour for the Midmark Ritter M11 (2nd-gen) Ultraclave Autoclave with a no-obligation diagnostic evaluation upon arrival. Don’t like the quote? We’ll send it back.

Free Return Shipping

We’ll send back your device for free regardless of its size or weight, anywhere in the world.

30-Day Guarantee

Acme guarantees service for no fewer than 30 days from return service delivery. Some device may qualify for extended warranty options.

Nearest Facility to You

Acme Revival Denver
2016 W Hamilton Place
Sheridan, CO 80110
Capacity: up to 150 tickets per week
Specialities: Medical, Industrial, Electro-Mechanical

Receiving Hours:
  • Monday 9:00 a.m. - 5:00 p.m.
  • Tuesday 9:00 a.m. - 5:00 p.m.
  • Wednesday 9:00 a.m. - 5:00 p.m.
  • Thursday 9:00 a.m. - 5:00 p.m.
  • Friday 9:00 a.m. - 5:00 p.m.
  • Saturday Closed
  • Sunday Closed

Join thousands of the world's largest companies that repair with Acme.