Topcon CT-80 non-contact Tonometer

Defect: NG- / NG+ is displaying on screen after air check

[/vc_column_text][vc_empty_space height=”65″][vc_tta_tabs active_section=”2″][vc_tta_section title=”Possible Cause” tab_id=”1615096150229-79543812-30aa”][vc_empty_space height=”35″][vc_column_text]The internal pump has failed, is clogged, or is dirty- preventing fluid movement of the pump to force air through the nozzle.[/vc_column_text][vc_empty_space height=”195″][/vc_tta_section][vc_tta_section title=”Solution” tab_id=”1615096150253-52387a85-51bb”][vc_empty_space height=”35″][vc_column_text]Attempt air-check procedures five times consecutively. If problem persists, the pump will most likely need to be replaced.

If above solution does not resolve defect, the device will most likely require service or repair.[/vc_column_text][vc_btn title=”Begin a Repair” link=”url:https%3A%2F%2Facmerevival.com%2Frepair-reuse-ui%2F|||”][vc_empty_space height=”85″][/vc_tta_section][/vc_tta_tabs][/vc_column][vc_column width=”1/6″][/vc_column][/vc_row][vc_row full_width=”stretch_row_content_no_spaces” kd_image_overlay_color=”#353535″ kd_fixed_background=”true” kd_image_overlay=”true”][vc_column][vc_empty_space height=”60″][/vc_column][/vc_row]